Laboratory Assessment Practice Report

   

Added on  2022-09-16

8 Pages1537 Words18 Views
Disease and Disorders
 | 
 | 
 | 
Laboratory Assessment Practice
University
Assessment
Name
Date
Laboratory Assessment Practice Report_1

Question 1 Assessment
06/06/2019
07/06/2019
08/06/2019
09/06/2019
10/06/2019
11/06/2019
12/06/2019
13/06/2019
14/06/2019
15/06/2019
16/06/2019
17/06/2019
18/06/2019
19/06/2019
20/06/2019
21/06/2019
22/06/2019
23/06/2019
24/06/2019
25/06/2019
2
2.05
2.1
2.15
2.2
2.25
2.3
2.35
2.4 UCL 2.345
CL 2.200
LCL 2.055
Calcium assay (mmol/L) Control 1
6/6/2019 - 6/25/2019
Calcium assay (mmol/L) Control 1
Figure 1 Calcium Assay control 1
06/06/2019
07/06/2019
08/06/2019
09/06/2019
10/06/2019
11/06/2019
12/06/2019
13/06/2019
14/06/2019
15/06/2019
16/06/2019
17/06/2019
18/06/2019
19/06/2019
20/06/2019
21/06/2019
22/06/2019
23/06/2019
24/06/2019
25/06/2019
2.35
2.55
2.75
2.95
3.15
3.35
3.55
3.75
3.95
4.15
4.35
UCL 4.066
CL 3.332
LCL 2.598
Calcium assay (mmol/L) Control 2
6/6/2019 - 6/25/2019
Calcium assay (mmol/L) Control 2
Figure 2 Calcium Assay Control 2
Laboratory Assessment Practice Report_2

Question 2
Accuracy and precision
06/06/201907/06/201908/06/201909/06/201910/06/201911/06/201912/06/201913/06/201914/06/201915/06/201916/06/201917/06/201918/06/201919/06/201920/06/201921/06/201922/06/201923/06/201924/06/201925/06/2019
0.00
0.50
1.00
1.50
2.00
2.50
3.00
3.50
4.00
CL 1.132
UCL 3.698
Calcium assay (mmol L XbarR Calcium
assay (mmol/L) Control 1-2 - R Chart
6/6/2019 - 6/25/2019
Range
Figure 3 Precision Control Chart
Quality control is an important parameter in detecting, reducing and correcting
inherent deficiencies in the laboratory internal process before the results are
relapsed. Quality control is essential in assessing the reliability of an essay.
Further quality control is essential in measuring precision and how well the test
is able to reproduce the same result over time under the operating conditions.
Quality control assessment is undertaken at the beginning of the shift. It is
essential for developing this assessment to ensure there is a minimal error of
reporting while no excessive repetitive analysis being done. Properties for quality
control entail same viscosity, turbidity, composition and colour. Further, they
need to be stable for a long period of time.
The above calcium assay figure 1 displays R4s rule indicating a shift in SD greater
than 2 demonstrating a random error while figure 2 display 22s rule
demonstrating systematic errors. The consecutive runs fall outside the 2s limit
values. This alters the accuracy of the assay is not detected what should detect.
Further, the precision aspect as observed in figure 3 demonstrates that the assay
is able to show the same effect of the result, there is accuracy in the manner in
which the assay is producing the results over time. Hence the essay