logo

Performance of Electrocardiogram

Measure the parameters of an ECG recorder using an oscilloscope and signal generator, and record test traces from an ECG waveform simulator.

7 Pages1053 Words87 Views
   

Added on  2023-06-04

About This Document

This article discusses the performance of electrocardiogram and its parameters. It covers the sensitivity of the recorder, circuit diagram, calculations, and more. The article also explains the role of low pass filter in removing unwanted signals and guaranteeing efficiency and effectiveness of the results obtained.

Performance of Electrocardiogram

Measure the parameters of an ECG recorder using an oscilloscope and signal generator, and record test traces from an ECG waveform simulator.

   Added on 2023-06-04

ShareRelated Documents
Performance of electrocardiogram 1
Performance of electrocardiogram.
Name
Institution
Date
Performance of Electrocardiogram_1
Performance of electrocardiogram 2
Introduction
Electrocardiogram is defined as the activity that involves the use of electrodes placed on
the human skin to record the heart’s electrical activities over a given time period (Rodriguez,
2017). It is so far the commonly practiced form of measuring the physiological potential in
humans. In this practical the electrocardiogram instrument to be used is an oscilloscope. An
oscilloscope is an electronic test instrument that instantly displays and analyses electronic
waveforms by drawing a two dimensional graph of the signal voltage against time (Johnson,
Brush, & McTigue, 2016). The oscilloscope in this practical will be used together with a signal
generator in measuring some crucial electrocardiogram recorder parameters. A wave form simulator
will also be used to obtain test traces that will be recorded and analyzed.
Methods
Apparatus
ECG recorder
Oscilloscopes
Signal generator
Potential divider
Crocodile clip lead
ECG waveform simulator
Method Procedure
i. The output of the generator was set to 10Hz sine wave while the sensitivity of the recorder
set to 1cm/Mv
ii. The input voltage of the signal generator was adjusted to give a deflection of 30 to 35 units
for every change in frequency of input signals.
iii. The apparatus were arranged as shown in figure 1 below
iv. All the lead-1 position which uses yellow and red terminus were measured
Performance of Electrocardiogram_2
Performance of electrocardiogram 3
v. The black terminus was connected to the earth of the signal generator side
vi. Peak-to-peak pen deflection of the recorder for peak-to-peak voltage of the signal generator
was recorded for the first 10 points
vii. Peak-to-peak pen deflection of the recorder for varying frequencies from 1 to 100 for a
constant input voltage was recorded.
viii. The results were used in plotting a graph.
Circuit diagram
Figure 1
Discussion on the arrangement of the resistors across the potential divider
The output voltage required guides on the arrangement of the resistors across the
potential divider. From the resistor values given on the potential divider and the required value
of output voltage for a given input voltage, equation Vout = Vin R2/(R1+R2) is used to locate the
resistors to give the required output voltage.
Performance of Electrocardiogram_3

End of preview

Want to access all the pages? Upload your documents or become a member.